| Model Number |
PFSD01008 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/27/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during navio tka procedure, when drilling bone pin, the surgeon realized it was bent.The procedure was completed without delay with an equipment swap.No patient harm or additional complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the navio bone pin, part pfsd01008 used for treatment was returned for evaluation.A relationship between the reported event and the device was confirmed.The reported problem was visually confirmed.The part is bent opposite the point.A functional evaluation could not be performed due to broken/damaged parts.While all products meet required manufacturing specifications prior to release a serial number or lot number is required to link the device to a dhr or nc investigation.A complaint history review for similar reported/confirmed complaints has identified prior events.The most likely cause of this event is component failure/damage due to the use of excessive force.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.
|
| |
|