Model Number 801763 |
Device Problem
Loss of Power (1475)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/21/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Per the field service representative (fsr), the perfusionist found a breaker that was tripped on the rear of the hlm prior to starting his case.The fsr could not duplicate the reported complaint.The unit operated to the manufacturer's specifications.
|
|
Event Description
|
It was reported that the heart lung machine (hlm) lost power.No other details regarding the nature of this event were provided.
|
|
Manufacturer Narrative
|
Per fsr, the facility had power interruption the night before the event occurred.
|
|
Event Description
|
Additional information received that there was no adverse consequences to the patient.
|
|
Manufacturer Narrative
|
Per data log analysis, on (b)(6) 2020 the system is powered up on battery backup at 06:27:16 am.At 08:35:44 am the system switches to alternating current (ac) power.It is possible the circuit breaker was tripped and then reset but there is no way to tell for sure from the log.The log does confirm ac power was lost.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|