Model Number 8637-40
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/10/2020
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: product id: 8709, serial#: (b)(4), implanted: (b)(6) 2007, explanted: (b)(6) 2020, product type: catheter.Other relevant device(s) are: product id: 8709, serial/lot #: (b)(4), ubd: 06-mar-2009, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from the healthcare provider via a device manufacturer representative regarding a patient receiving morphine (10 mg/ml at 2 mg/day) via an implantable infusion pump.It was reported that during a scheduled pump replacement the hcp determined the old catheter was not working.He was unable to get flow through the side port and unable to aspirate from the catheter directly.The old catheter was explanted and a new ascenda catheter was implanted in its place.The hcp discarded the old catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
david gustafson
7000 central avenue ne rcw215
minneapolis, MN 55432
7635149628
MDR Report Key10836340
MDR Text Key216132584
Report Number3004209178-2020-19945
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00643169100824
UDI-Public00643169100824
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 11/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/14/2015
Device Model Number8637-40
Device Catalogue Number8637-40
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/10/2020
Initial Date FDA Received11/13/2020
Date Device Manufactured06/27/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age59 YR
-
-