• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 3.2MM DRILL BIT/QC/145MM; INSTR,SURGICAL,ORTHOPEDIC,AC- POWERED MOTOR/ACCESS & ATTACH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 3.2MM DRILL BIT/QC/145MM; INSTR,SURGICAL,ORTHOPEDIC,AC- POWERED MOTOR/ACCESS & ATTACH Back to Search Results
Model Number 310.31
Medical Device Problem Codes Break (1069); Entrapment of Device (1212)
Health Effect - Clinical Codes Injury (2348); Foreign Body In Patient (2687)
Date of Event 10/18/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional procode: gff, gfa, hsz.Complainant part is not expected to be returned for manufacturer review/ investigation.Reporter is a j&j employee.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/ or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2020 during orif distal femur procedure, the drill bit broke while drilling.The surgeon was hitting a lot of metal and bone cement.The tip of the drill bit was left in the patient.There was no surgical delay.The procedure was successfully completed.Patient outcome was unknown.This report is for one (1) 3.2mm drill bit/qc/145mm.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.B5: additional event description updated.H6: codes updated.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Additional event description.Concomitant device reported: unknown bone cement (part #: unknown, lot #: unknown, quantity: 1).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3.2MM DRILL BIT/QC/145MM
Common Device Name
INSTR,SURGICAL,ORTHOPEDIC,AC- POWERED MOTOR/ACCESS & ATTACH
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key10837264
Report Number2939274-2020-05149
Device Sequence Number11712197
Product Code HWE
UDI-Device Identifier10886982186478
UDI-Public(01)10886982186478
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number310.31
Device Catalogue Number310.31
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/18/2020
Supplement Date Received by Manufacturer11/13/2020
Initial Report FDA Received Date11/13/2020
Supplement Report FDA Received Date12/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age86 YR
-
-