MAUDE Adverse Event Report: DEPUY MITEK LLC US 4580 FMS DUO+ PUMP/SHAVER COMBO -NS; DISTENSION UNIT, FLUID, ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 4580 FMS DUO+ PUMP/SHAVER COMBO -NS; DISTENSION UNIT, FLUID, ARTHROSCOPIC Back to Search Results
Model Number 284580
Device Problem Increase in Pressure (1491)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4).The product upon which this medwatch is based has been received, however, the product evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
 
Event Description
It was reported by the affiliate in (b)(6) that the pump/shaver device had a substantial level of pressure to 900 during the shaver suction test, and that it passed the maximum level of 875.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in the surgical procedure, or if a spare device was available for use.There were no reports of any injuries, medical intervention, or prolonged hospitalization.The exact date of the event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: the complaint device was received at the service center and evaluated.It was reported that the item had a substantial level of 900 at the shaver suction test led 4.The maximum level should be 875 and the grooving tool was worn out.Per service reports, this complaint can be confirmed.It was found during evaluation that the two wings were missing.Further, unit was not calibrated correctly.The missing components were renewed, defective material exchanged, pressure mechanism re-adjusted, mechanical upgrade performed, and unit was cleaned resolve the issues.After repair, the device was found to be working according to the specifications.The missing wings are most likely a result of user mishandling or improper maintenance.Also, user error is the most likely root cause of the incorrect calibration of the device.However, given the information provided, we cannot discern a definitive root cause of the other defective parts of the device.The missing wings and incorrect calibration would have caused the customer to experience the reported problem.The service history has been reviewed in lieu of the device history record for this device since it was previously serviced.The device was last serviced on 11/17/2016 and passed all functional testing before being returned to the customer.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4580 FMS DUO+ PUMP/SHAVER COMBO -NS
Type of Device
DISTENSION UNIT, FLUID, ARTHROSCOPIC
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key10837482
MDR Text Key216178118
Report Number1221934-2020-03473
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705020607
UDI-Public10886705020607
Combination Product (y/n)N
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/13/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number284580
Device Catalogue Number284580
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2020
Date Manufacturer Received11/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-