| Model Number |
1365-36-220 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); No Code Available (3191)
|
| Date of Event |
06/25/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Clinical notification received for revision of right total hip to address infection.Date of implantation: (b)(6) 2018, date of revision: (b)(6) 2019, (right hip).Treatment: revision of cup, 2 screws, hole eliminator, liner, stem, proximal sleeve, and head--placement of antibiotic spacer.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|