Model Number ADVIA 2120I HEMATOLOGY SYSTEM WITH DUAL ASPIRATE AUTOSAMPLER
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/21/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer contacted a siemens customer care center (ccc).An onsite siemens customer service engineer (cse) serviced the advia 2120i hematology system with dual aspirate autosampler and noted that the red blood cell (rbc) rate was inconsistent.Platelets are counted in the rbc channel, and therefore if the rate is inconsistent for rbc, it will be inconsistent for platelets as well since they are counted at the same time.The cse backflushed the flowcell which improved the rbc rate.The cse also inspected the rbc optics and adjusted the pulse width to be within specification.Quality controls (qc) recovered in range after the service intervention and the system was operational.The cause of the falsely low platelet result was the dirty flowcell and/or the rbc optics.The system is performing according to specifications.No further evaluation of this device is required.
 
Event or Problem Description
A discordant, falsely low platelet result was obtained on a patient sample on an advia 2120i hematology system with dual aspirate autosampler.The discordant result was reported to the physician(s) and was questioned.The sample was repeated for platelet on an alternate advia 2120i hematology system, resulting higher.The higher result was reported, as the correct result, to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely low platelet result.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVIA 2120I HEMATOLOGY SYSTEM WITH DUAL ASPIRATE AUTOSAMPLER
Common Device Name
ADVIA 2120I HEMATOLOGY SYSTEM WITH DUAL ASPIRATE AUTOSAMPLER
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC
511 benedict ave
tarrytown NY 10591
MDR Report Key10838428
Report Number2432235-2020-00449
Device Sequence Number3274803
Product Code GKZ
UDI-Device Identifier00630414560045
UDI-Public00630414560045
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K102644
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberADVIA 2120I HEMATOLOGY SYSTEM WITH DUAL ASPIRATE AUTOSAMPLER
Device Catalogue Number10285573
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/24/2020
Initial Report FDA Received Date11/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-