• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ILS 33MM, STRAIGHT; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ILS 33MM, STRAIGHT; STAPLE, IMPLANTABLE Back to Search Results
Model Number SDH33A
Medical Device Problem Code Failure to Form Staple (2579)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/25/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch #: unknown.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.
 
Event or Problem Description
It was reported that during the resection of low rectal cancer surgery, when severing rectum and colon tissue, after fired, noted staples malformed with irregular shape (with straight leg).Changed the new one to complete surgery.There was no patient consequence reported.No additional information can be provided.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 12/09/2020.D4: batch # r57157.Device analysis: the analysis results found that the sdh33a device arrived with no apparent damage.The breakaway washer was present and uncut, the device was fully loaded with staples, indicating that the device had not being fired.The device was tested for functionality with the returned washer and it fired and formed all the staples as well as completely cut the test media and the breakaway washer without incident.The staple line was complete, and the staples meet the staple form release criteria.Although there is no direct evidence of use error, the instructions for use do contain the following caution: if excessive force is required to close the device, this may indicate there is too much tissue or thickened tissue in the device.Attempting to fire the instrument in this condition may result in poor staple line integrity with possible leakage or disruption.In addition, instrument damage or failure may result.The event described could not be confirmed as the device performed without any difficulties.As part of our quality process, the manufacturing records of this batch was reviewed, and the manufacturing standards were met prior to the release of this batch.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ILS 33MM, STRAIGHT
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key10838737
Report Number3005075853-2020-06000
Device Sequence Number3501533
Product Code GDW
UDI-Device Identifier10705036003540
UDI-Public10705036003540
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K983536
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/10/2022
Device Model NumberSDH33A
Device Catalogue NumberSDH33A
Device Lot NumberR4022L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2020
Initial Date Received by Manufacturer 10/25/2020
Supplement Date Received by Manufacturer11/19/2020
Initial Report FDA Received Date11/13/2020
Supplement Report FDA Received Date12/09/2020
Patient Sequence Number1
-
-