Catalog Number 2C8858 |
Device Problems
Fluid/Blood Leak (1250); Material Split, Cut or Torn (4008)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 10/13/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device manufacturer address 1: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the tubing of a paclitaxel set had a small cut leading to a medication leak.The set was used with an unspecified pump.This issue was identified prior to patient infusion.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
The actual device was not available; however, a companion sample was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed including clear passage and pressure testing; and the device performed according to product specifications.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|