BAXTER HEALTHCARE CORPORATION CLEARLINK SOLUTION ADMINISTRATION SETS; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2C8858
Device Problems Fluid/Blood Leak (1250); Material Split, Cut or Torn (4008)
Patient Problem No Patient Involvement (2645)
Event Date 10/13/2020
Event Type  malfunction  
Manufacturer Narrative
Device manufacturer address 1: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the tubing of a paclitaxel set had a small cut leading to a medication leak.The set was used with an unspecified pump.This issue was identified prior to patient infusion.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The actual device was not available; however, a companion sample was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed including clear passage and pressure testing; and the device performed according to product specifications.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK SOLUTION ADMINISTRATION SETS
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key10839242
MDR Text Key216359311
Report Number1416980-2020-07114
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00085412071688
UDI-Public(01)00085412071688
Combination Product (y/n)Y
Reporter Country CodeCA
PMA/PMN Number
K160007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2C8858
Device Lot NumberR20D04059
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/21/2020
Initial Date FDA Received11/13/2020
Supplement Dates Manufacturer Received12/15/2020
Supplement Dates FDA Received12/22/2020
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/05/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-