COOK INC FLEXOR ANSEL GUIDING SHEATH; DYB INTRODUCER, CATHETER Back to Search Results
Model Number G44155
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 10/21/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Occupation: title: rtr.(b)(4).A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
As reported, at the completion of a procedure involving the leg, a flexor ansel guiding sheath separated at the hub upon removal of the device from the right groin.Unspecified catheters and another manufacturer's atherectomy device were used through the sheath, which was in place for the entire case, approximately two hours.The anatomy was scarred, and resistance was reported upon both insertion and removal of the sheath.The hub and sheath were both used to pull the device from the patient over an unknown wire guide.The dilator was not in place upon removal of the device.As the device was removed, the "clear portion that connects to the main hub" broke.The separated portion was removed with hemostats.The procedure was successfully completed at the time of the event.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Event or Problem Description
No new patient or event information to report.
 
Additional Manufacturer Narrative
Description of event: as reported, at the completion of a procedure involving the leg, a flexor ansel guiding sheath separated at the hub upon removal of the device from the right groin.Unspecified catheters and another manufacturer's atherectomy device were used through the sheath, which was in place for the entire case, approximately two hours.The anatomy was scarred, and resistance was reported upon both insertion and removal of the sheath.The hub and sheath were both used to pull the device from the patient over an unknown wire guide.The dilator was not in place upon removal of the device.As the device was removed, the "clear portion that connects to the main hub" broke.The separated portion was removed with hemostats.The procedure was successfully completed at the time of the event.Investigation ¿ evaluation.A visual inspection of the returned device was conducted.A document based investigation was also performed including a review of complaint history, device history record, drawings, the instructions for use, manufacturing instructions, and quality control data.A review of the device history record found no non-conformances related to the reported failure mode.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in field.A review of complaint history records shows no other complaints associated with the complaint device lot.The instructions for use (ifu), provides the following information to the user related to the reported failure mode: "warnings: if resistance is encountered during advancement of flexor sheath, assess cause of resistance and consider dilation of any restriction identified or consider alternate treatment strategy.If flexor sheath is advanced through resistance, force to remove the sheath will be higher, increasing the risk of sheath material or hub separation upon withdrawal.Reinsertion of dilator prior to removal of flexor sheath increases the strength of the sheath and lessens the risk of device separation.If resistance is anticipated or encountered during withdrawal of flexor sheath, consider carefully reinserting the dilator prior to continuing removal.Precautions: in order to ensure device compatibility, choose a sheath size large enough to accommodate the maximum outer diameter of any devices that will be placed through the sheath.All interventional or diagnostic instruments used with this product should move freely through the valve and sheath to avoid damage.Instructions for use: sheath introduction.1.Ensure that the inner diameter (id) of the sheath is appropriate for the maximum diameter of the instruments to be introduced.2.Using the side-arm of the valve, flush the sheath by filling the sheath assembly completely with heparinized saline.3.Flush the dilator with heparinized saline.How supplied: upon removal from package, inspect the product to ensure no damage has occurred." a review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was visually/functionally inspected by quality control and no related gaps in production or processing controls were noted.Based on the provided evidence and the completed investigation, cook has concluded patient anatomy and user error may have contributed to this incident.Per the quality engineering risk assessment, no further action is warranted.Cook will continue to monitor this device via the complaints database for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXOR ANSEL GUIDING SHEATH
Common Device Name
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key10839897
Report Number1820334-2020-02092
Device Sequence Number4142135
Product Code DYB
UDI-Device Identifier00827002441552
UDI-Public(01)00827002441552(17)230521(10)13205320
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K142829
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/21/2023
Device Model NumberG44155
Device Catalogue NumberKCFW-7.0-18/38-45-RB-ANL0-HC
Device Lot Number13205320
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/17/2020
Initial Date Received by Manufacturer 11/05/2020
Supplement Date Received by Manufacturer02/10/2021
Initial Report FDA Received Date11/13/2020
Supplement Report FDA Received Date02/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
JETSTREAM ATHERECTOMY DEVICE
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient Weight74
-
-