PENUMBRA, INC. INDIGO SYSTEM ASPIRATION CATHETER 8; DXE Back to Search Results
Model Number CAT8XTORQ115KIT-B
Medical Device Problem Codes Material Puncture/Hole (1504); Material Twisted/Bent (2981)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/21/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure in the femoral popliteal artery using an indigo system aspiration catheter 8 (cat8), a penumbra engine (engine), a guidewire and non-penumbra sheath.It was noted that there was stenosis in the target vessel.During the procedure, the physician advanced the cat8 over the guidewire, through the sheath and into the target vessel to perform aspiration.Towards the end of the procedure, after completing three passes using the cat8, the physician noticed full aspiration could not be achieved with the engine.Upon examining the cat8, the physician observed that the proximal end of the cat8 was twisted and had a small hole where it was being torqued.Therefore, the cat8 was removed.The procedure was completed using an indigo system aspiration catheter 6 (cat6) and the same sheath.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INDIGO SYSTEM ASPIRATION CATHETER 8
Common Device Name
DXE
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key10839933
Report Number3005168196-2020-01992
Device Sequence Number16962686
Product Code DXE
UDI-Device Identifier00815948022485
UDI-Public00815948022485
Combination Product (Y/N)Y
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K160533
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/21/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/15/2023
Device Model NumberCAT8XTORQ115KIT-B
Device Catalogue NumberCAT8XTORQ115KIT
Device Lot NumberH10624
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/21/2020
Initial Report FDA Received Date11/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age52 YR
Patient Weight127
Date Report Sent to FDA01/01/2005
-
-