• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. RIM SPEED PIN 45MM STERILE; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. RIM SPEED PIN 45MM STERILE; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 74013471
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/23/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during total knee arthroplasty the speedpin head worn away, making it impossible to remove from the cutting block with a drill.A pin puller was used to twist it out.There was a delay reported of less than or equal to 30 minutes.The procedure was finished using a smith and nephew backup device.
 
Additional Manufacturer Narrative
Results of investigation: the device, used in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A medical investigation was conducted and confirms this case reports the head of the speedpin wore away, making it necessary to use a pin puller to twist it out of the cutting block, as it could not be removed with the drill.Per email communication, there was no patient injury, and the procedure was completed using a backup device, with minimal delay.Therefore, since no patient harm is alleged, no further clinical assessment is necessary.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.Damage from misuse or rough handling are likely probable causes of the reported event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RIM SPEED PIN 45MM STERILE
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10840403
Report Number1020279-2020-06445
Device Sequence Number3799626
Product Code HSX
UDI-Device Identifier00885556069981
UDI-Public00885556069981
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K113038
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number74013471
Device Catalogue Number74013471
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/26/2020
Supplement Date Received by Manufacturer03/25/2021
Initial Report FDA Received Date11/13/2020
Supplement Report FDA Received Date04/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-