| Model Number |
AC*T DIFF 2 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/20/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The field service engineer (fse) was on site and confirmed what the customer reported.To resolve the issue the fse replaced the rbc bath after which plt counts were normal resolving the reported issue.The fse ran numerous start-ups with no further high plt backgrounds or noise spikes.Bec internal identifier: (b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported recovering erroneous high platelet (plt) results for a single patient sample run on the act diff 2 analyzer after onsite service.There was no report of erroneous patient results reported outside the laboratory.There was no report of impact to patient treatment.
|
| |
|
Search Alerts/Recalls
|