Model Number 254730
Medical Device Problem Codes Degraded (1153); Dull, Blunt (2407)
Health Effect - Clinical Codes Tissue Damage (2104); Not Applicable (3189); Unspecified Tissue Injury (4559)
Date of Event 10/31/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Udi: (b)(4).
 
Event or Problem Description
It was reported by affiliate via complaint submission tool that while harvesting the semitendinosus tendon(tendon harv closed stripper ea), the tendon tear prematurely so the tendon was too short to use as a replacement for the acl reconstruction.The surgeon complain that a part of this problem was that the stripper was not sharp enough.Surgeon had to harvest a part of the tractus iliotibialis.30 minutes delay.No patient consequences.The device is available for evaluation.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary =according to the information provided, it was reported that while harvesting the semitendinosus tendon, the tendon tear prematurely so the tendon was too short to use as a replacement for the acl reconstruction.The surgeon complain that a part of this problem was that the stripper was not sharp enough.Surgeon had to harvest a part of the tractus iliotibialis.The complaint device was received and evaluated.Visual observation reveals that the device is in normal use conditions, no structural damages could be found.The markings on the shaft are legible.The shaft is not bent.When reviewing at the distal tip of the device, it could be observed that the tip is dull.According with the visual inspection result, this complaint can be confirmed.This failure can be attributed to fair wear and tear of the device due to repeated use and sterilization cycles.A manufacturing record evaluation was performed for the finished device [19e01] number, and no non-conformances were identified.At this time, no corrective action is required, and no further action is warranted, as the device is fully functional.However, depuy synthes mitek will continue to track any related complaints within this device family to monitor the extent to which this complaint is observed in the field.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi:(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDON HARV CLOSED STRIPPER EA
Common Device Name
BONE TREPHINE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key10843366
Report Number1221934-2020-03490
Device Sequence Number14180157
Product Code LXH
UDI-Device Identifier10886705013739
UDI-Public10886705013739
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number254730
Device Catalogue Number254730
Device Lot Number19E01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2020
Supplement Date Received by Manufacturer12/07/2020
12/20/2020
Initial Report FDA Received Date11/16/2020
Supplement Report FDA Received Date12/08/2020
12/21/2020
Patient Sequence Number1
-
-