U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-20
Medical Device Problem Code Disconnection (1171)
Health Effect - Clinical Code Seroma (2069)
Date of Event 11/01/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products: product id: 8782, serial#:(b)(4), implanted: (b)(6) 2018, product type: catheter.Other relevant device(s) are: product id: 8782, serial/lot #: (b)(4), ubd: 02-nov-2018, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a healthcare provider via a company representative regarding a patient receiving an unknown drug conce ntrations reported as ¿5 and 2.5¿ and dose reported as ¿1.07 and 12¿ via an implantable pump.It was reported that the patient had their pump in their back and the patient was hit in the back with an object and the pocket filled with some fluid.Per the reporter it was unknown if the collection of fluid was a seroma.A revision was performed and the spinal piece came out of the collet.A piece of the catheter was trimmed and they changed the concentration to lower the dose since the patient wasn¿t getting any medication.A dye study was performed and the catheter was currently cleared.The reporter was guided on proper bridge bolus parameters for new catheter length.
 
Event or Problem Description
Additional information was received from a healthcareprofessional (hcp) via a company representative who reported that during the procedure, the catheter was revised by connecting the 2 existing pieces of the catheter with a new collet.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Common Device Name
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key10843674
Report Number3004209178-2020-20035
Device Sequence Number14180168
Product Code LKK
UDI-Device Identifier00643169984219
UDI-Public00643169984219
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P860004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/28/2020
Device Model Number8637-20
Device Catalogue Number8637-20
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/11/2020
Supplement Date Received by Manufacturer11/17/2020
Initial Report FDA Received Date11/16/2020
Supplement Report FDA Received Date11/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"SEE H10...."; "SEE H10...."
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
-
-