|
A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: analysis of the returned device identified crease flat, white particles in the shell, and bubbles in the gel observed after autoclave disinfection cycle.Weight measurement of the device identified device weight was within specification.Based on the device analysis the final assessment was no issues found related with the manufacturing process.The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii.
|