| Model Number |
45038 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/27/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was cancelled.An angiojet ultra 5000a console was used for a thrombectomy procedure.During the preparation, it was noted that the foot pedal was no longer working.No physical damage noted on the foot pedal.The procedure was cancelled.No patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was cancelled.An angiojet ultra 5000a console was used for a thrombectomy procedure.During the preparation, it was noted that the foot pedal was no longer working.No physical damage noted on the foot pedal.The procedure was cancelled.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: the ultra foot switch was received in bad condition with physical damages and defects observed.The ultra foot switch failed the functional test.There were bent pins found on a connector socket, therefore the foot switch not working was confirmed.
|
| |
|