| Model Number |
7010 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/27/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter facility name: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 2.5mm x 12mm quantum maverick balloon catheter was advanced for use; however, during procedure, the device was fractured.The procedure was completed with another of the same device.No patient complications were reported and the patient status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
E1 - initial reporter facility name: (b)(6).Device evaluation by mfr: returned product consisted of a quantum maverick balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was a complete separation at 49.3cm distal of the strain relief.There was blood in the guidewire lumen.The balloon was tightly folded.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed the reported broken shaft as the hypotube was separated.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 2.5mm x 12mm quantum maverick balloon catheter was advanced for use; however, during procedure, the device was fractured.The procedure was completed with another of the same device.No patient complications were reported and the patient status was stable.
|
| |
|