It was reported that during an arthroscopy the screw was turned away when the screw was turned in, it broke.The procedure was completed with a backup device and no significant delay or further complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
The reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection found one screw intact and the second is screw broken missing the bottom half.Based on the condition of the product material found during visual inspection it was determined that additional material testing is not required.A review of the raw materials, found that the storage requirements, material specifications, and material tests were appropriately documented.A review of the device records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated event.A review of risk management files found that the reported failure was documented appropriately.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.The complaint was confirmed.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device tip, attempted correction of a damaged device, or an inadvertent impact event.No containment or corrective actions are recommended at this time.
|