| Model Number |
7010 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/19/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The 80% stenosed target lesion was located in the mildly tortuous and moderately calcified left circumflex artery.A 2.5mm x 12mm quantum maverick balloon catheter was advanced from dilatation.However, during procedure, the catheter was fractured outside the interventional vessel, 64cm from the hub.The device was completely removed directly from the patient's body and the procedure was completed with another of same device.No patient complications were reported and the patient's status was fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of a quantum maverick balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There was a complete separation at 70.1cm distal of the strain relief.There was no distal portion of the device returned.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed the reported broken shaft as the hypotube was separated.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The 80% stenosed target lesion was located in the mildly tortuous and moderately calcified left circumflex artery.A 2.5mm x 12mm quantum maverick balloon catheter was advanced from dilatation.However, during procedure, the catheter was fractured outside the interventional vessel, 64cm from the hub.The device was completely removed directly from the patient's body and the procedure was completed with another of same device.No patient complications were reported and the patient's status was fine.
|
| |
|
Search Alerts/Recalls
|