MAUDE Adverse Event Report: ZIMMER BIOMET SPINE INC. CURVE VARIABLE INSERTER SHAFT; ZYSTON ARC INTERBODY SPACER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE INC. CURVE VARIABLE INSERTER SHAFT; ZYSTON ARC INTERBODY SPACER Back to Search Results
Catalog Number 14-533041
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 10/16/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
 
Event or Problem Description
It was reported that the tip of the zyston curve inserter's draw rod broke off inside a cage during surgery at the time of the final impaction.The cage with the tip was left implanted in the patient and bone was packed behind it.There were no immediate patient impacts reported.
 
Additional Manufacturer Narrative
The complaint is confirmed for one returned inserter shaft for the failure of tip fractured and remains inside the body.Visual inspection revealed the threaded tip has fractured off the inserter shaft.Medical records were not provided for review.Potential cause: the cause of the damage cannot be determined at this time since there is no information available regarding how the inserter was being used or handled at the time of the damage.Dhr review and related actions: per dhr review, the part was likely conforming when it left zimmer biomet control.No actions required.This event is not related to any current actions or recalls or product holds.Device use: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Event or Problem Description
It was reported that the tip of the zyston curve inserter's draw rod broke off inside a cage during surgery at the time of the final impaction.The cage with the tip was left implanted in the patient and bone was packed behind it.There were no immediate patient impacts reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CURVE VARIABLE INSERTER SHAFT
Common Device Name
ZYSTON ARC INTERBODY SPACER
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
MDR Report Key10848551
Report Number3012447612-2020-00627
Device Sequence Number19662424
Product Code MAX
UDI-Device Identifier00888480470182
UDI-Public(01)00888480470182(10)PT15D
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K110650
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number14-533041
Device Lot NumberPT15D
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 10/20/2020
Supplement Date Received by Manufacturer02/15/2021
Initial Report FDA Received Date11/16/2020
Supplement Report FDA Received Date02/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-