Model Number 71367958
Device Problem Break (1069)
Patient Problems No Consequences Or Impact To Patient (2199); No Information (3190); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/23/2020
Event Type  malfunction  
Event Description
It was reported that the r3 20 degrees 36mm trial 58mm is chipped and broken.It is unknown where the event occurs.
 
Manufacturer Narrative
The device, intended for use in treatment was returned for evaluation.Visual inspection of the returned device confirms that the device is chipped, damaged and has signs of wear and tear from use.The device was manufactured in 2018.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.A review of complaint history for the listed part revealed no prior complaints.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.The device is a reusable device that can be exposed to numerous surgeries and cleaning cycles.Probable causes that could contribute to the reported event, as reusable instruments are susceptible to damage from prolonged use and through misuse or rough handling.To avoid compromised performance or damage, it is recommended the device be inspected prior to and after each use and cleaning.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further actions are being taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 20DEG 36MM TRIAL 58MM
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10850003
MDR Text Key216562315
Report Number1020279-2020-06526
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010599902
UDI-Public03596010599902
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 11/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/17/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71367958
Device Catalogue Number71367958
Device Lot Number18DDS0004
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2020
Date Manufacturer Received11/19/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-