| Model Number |
421033 |
| Medical Device Problem Codes |
Material Disintegration (1177); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/26/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device 2 of 2.Based on the available information, this event is deemed to be a reportable malfunction.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that for 2 out of 10 wafers from 1 market unit, when the end user attempted to mold, the center of the moldable material was shredding and coming apart.This occurred prior to use, but she used them anyway as she had no other wafers on hand.There was no harm reported.A photograph depicting the issue was received from the complainant.
|
| |
|
Additional Manufacturer Narrative
|
|
Returned sample evaluation: photograph associated with this case were received in which the reported defect can be observed.No unused return sample was received for evaluation.Conclusion summary of the related event: based on the results of the preliminary investigation, the batch record review was performed and no discrepancies were found.In addition, the failure modes reported could not be replicated.Furthermore, through the process investigation it was concluded that the reported issues can be associated to the mixing process.The root cause investigation tw#(b)(4) was generated in which the causes and actions identified are directly related with the problems reported by the customers.For all the explanation above, a root cause investigation (rci) is not necessary.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 1049092 manufacturing site: 9618003.
|
| |
|
Event or Problem Description
|
|
To date no additional patient or event details have been received.
|
| |
|