MAUDE Adverse Event Report: LEONHARD LANG GMBH STABLE BASE; ECG ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LEONHARD LANG GMBH STABLE BASE; ECG ELECTRODE Back to Search Results
Model Number T-601
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Irritation (1941); Itching Sensation (1943); Rash (2033); Burning Sensation (2146)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Retained samples of the concerned lot/batch (191029-0155, 191104-0157 and 200130-0259) of model sbt601 have been inspected visually.Mechanical tests were performed on 3 retained samples each.All tested samples were found to perform within limits.No faults could be detected.No information on the patient, skin type, state of skin, whether any medication was being taken, which might have a skin weakening effect, duration of use and details of the use was provided.We have requested further information several times but none was made available.The initial reporter was specifying "we continue to encourage the customer to provide this information.However we can't force them to do so.Our customer is a very large healthcare provider in the united states and they have a general system for patient complaints.They will not develop a special tracking system and process for (b)(6)." we therefore consider the investigation closed.
 
Event or Problem Description
On (b)(6) 2020, we have been informed about 14 incidents with ecg electrodes.Monitoring ecg electrodes (model sbt601) had been used with body guardian one and body guardian verite one devices.The initial reporter has attached for each patient an incident summary.In total, there were 14 reports covering a period from july 02 to august 17.No medical intervention was necessary for 13 incidents.For 1 incident a medical intervention was needed to treat the injury but no further details were given.Before applying the ecg electrodes, the skin of the patient requiring intervention was prepared using water and soap.He developed open sores and bleeding.The patient paused the study to wait for the skin to heal.A medical intervention was necessary.All other patients were prepared with different methods using water and soap, alcohol, alcohol wipes or saline wipes before applying the ecg electrodes.The patients were experiencing the "redness, itching, rash, blisters, welt, burning sensation on skin, peeling skin, burnt skin, bleeding, open sores." no further information was provided.
 
Event or Problem Description
On october 19th, 2020, we have been informed about 14 incidents with ecg electrodes.Monitoring ecg electrodes (model sbt601) had been used with bodyguardian one and bodyguardian verite one devices.The initial reporter has attached for each patient an incident summary.In total, there were 14 reports covering a period from (b)(6).No medical intervention was necessary for 13 incidents.For 1 incident a medical intervention was needed to treat the injury but no further details were given.Before applying the ecg electrodes, the skin of the patient requiring intervention was prepared using water and soap.He developed open sores and bleeding.The patient paused the study to wait for the skin to heal.A medical intervention was necessary.All other patients were prepared with different methods using water and soap, alcohol, alcohol wipes or saline wipes before applying the ecg electrodes.The patients were experiencing the "redness, itching, rash, blisters, welt, burning sensation on skin, peeling skin, burnt skin, bleeding, open sores." no further information was provided.
 
Additional Manufacturer Narrative
Retained samples of the concerned lot/batch (191029-0155, 191104-0157 and 200130-0259) of model sbt601 have been inspected visually.Mechanical tests were performed on 3 retained samples each.All tested samples were found to perform within limits.No faults could be detected.No information on the patient, skin type, state of skin, whether any medication was being taken, which might have a skin weakening effect, duration of use and details of the use was provided.We have requested further information several times but none was made available.The initial reporter stated that no further information has been made available despite repeated requests.We therefore consider the investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STABLE BASE
Common Device Name
ECG ELECTRODE
Manufacturer (Section D)
LEONHARD LANG GMBH
archenweg 56
innsbruck, 6020
AU  6020
MDR Report Key10850224
Report Number8020045-2020-00028
Device Sequence Number2322264
Product Code DRX
UDI-Device Identifier10861779000274
UDI-Public(01)10861779000274
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K023503
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Expiration Date01/30/2023
Device Model NumberT-601
Device Catalogue NumberSBT601
Device Lot Number200130-0259
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/19/2020
Supplement Date Received by Manufacturer10/19/2020
Initial Report FDA Received Date11/17/2020
Supplement Report FDA Received Date02/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-