U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII OVAL RESURFACING PAT 29MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII OVAL RESURFACING PAT 29MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, Back to Search Results
Model Number 71421029
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Injury (2348)
Date of Event 10/23/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient received a revision left total knee replacement on (b)(6) 2020.Patient has complained of anterior knee pain increasing in severity.A revision surgery was performed on (b)(6) 2020 and the patella bone was resected and new resurfacing patella implanted.Surgeon believed this was 'overstuffed' from x-ray images and patient symptoms.Patient outcome is unknown.
 
Additional Manufacturer Narrative
It was reported that a revision surgery was performed as patient has complained of anterior knee pain increasing in severity.The patella bone was resected and new resurfacing patella implanted.The affected complaint device, used in treatment, was not returned for evaluation.Therefore a product analysis could not be performed.Our investigation including a review of the manufacturing records for the listed batch did not reveal any deviation from the standard manufacturing processes.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of risk management files and the instructions for use found that the reported failure was documented appropriately.Based on this investigation, the need for corrective action is not indicated.A clinical evaluation noted that based on the x-rays/information provided, the reported ¿overstuffing¿ and laterally riding patella were likely the root cause of the reported event.The patient impact beyond the reported increasing pain and subsequent revision procedure could not be determined, as the patient outcome remains unknown.No further investigation is warranted for this complaint; however smith and nephew will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII OVAL RESURFACING PAT 29MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED,
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10850391
Report Number1020279-2020-06527
Device Sequence Number3630053
Product Code JWH
UDI-Device Identifier03596010485052
UDI-Public03596010485052
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K951987
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number71421029
Device Catalogue Number71421029
Device Lot Number19GM04625
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 10/23/2020
Supplement Date Received by Manufacturer10/23/2020
Initial Report FDA Received Date11/17/2020
Supplement Report FDA Received Date02/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-