SMITH & NEPHEW, INC. R3 LNR MPACTOR HD II 28MM; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Model Number 71366428
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 10/23/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during inspection the head of the r3 lnr mpactor hd ii 28mm was found broken.No case related, therefore no patient involved.
 
Additional Manufacturer Narrative
H3, h6: the device, intended for use in treatment was returned for evaluation.A visual inspection of the returned device confirmed the stated failure mode.The head of the device is cracked, rendering it inoperable.The device show signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 LNR MPACTOR HD II 28MM
Common Device Name
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10850462
Report Number1020279-2020-06535
Device Sequence Number14360013
Product Code JDG
UDI-Device Identifier03596010656056
UDI-Public03596010656056
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71366428
Device Catalogue Number71366428
Device Lot Number13JSY0003
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2020
Initial Date Received by Manufacturer 10/23/2020
Supplement Date Received by Manufacturer10/01/2021
Initial Report FDA Received Date11/17/2020
Supplement Report FDA Received Date10/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-