BOSTON SCIENTIFIC CORPORATION ENCORE 26; SYRINGE, BALLOON INFLATION Back to Search Results
Model Number M0067101140
Medical Device Problem Codes Mechanical Problem (1384); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/28/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
User facility/importer report# (b)(4).(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
Note: this report pertains to an encore inflation device and a nephromax dilatation balloon catheter used in the same patient and procedure.It was reported to boston scientific corporation that an encore inflation device and a nephromax dilatation balloon catheter were used in the kidney during a percutaneous nephrolithotomy (pcnl) procedure performed on (b)(6) 2020.According to the complainant, during the procedure, the balloon burst when inflated to 16 psi.Subsequently, the inflation medium of the balloon was found filling the retroperitoneum of the patient which made the image of the kidney under fluoroscopy too challenging to see.Reportedly, there are two wires in the kidney at that time.The encore inflation device was noted hard to use.The procedure was not completed due to this event, and the patient had to be brought back to the hospital at a later date for another procedure.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENCORE 26
Common Device Name
SYRINGE, BALLOON INFLATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key10850696
Report Number3005099803-2020-05510
Device Sequence Number2419373
Product Code MAV
UDI-Device Identifier08714729755814
UDI-Public08714729755814
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K955869
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/19/2023
Device Model NumberM0067101140
Device Catalogue Number710-114
Device Lot Number0025522830
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/23/2020
Initial Report FDA Received Date11/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age57 YR
-
-