Model Number SERVO-I
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/06/2020
Event Type  malfunction  
Manufacturer Narrative
The ventilator was investigated by our field service engineer (fse).The reported failure was confirmed and the user interface holder was replaced.The ventilator was returned to clinical use.No parts were returned for investigation.The consequence of this kind of mechanical damage is that the user interface gets detached and, in a worst case scenario, falls off the ventilator carrier.Our conclusion into this matter is that the user interface has been exposed to a mechanical force greater than it is designed to sustain.The ventilator system was successfully tested for mechanical strength, with a peak acceleration of 15 g, pulse duration 6 ms, and total number of 1000 impacts.
 
Event Description
It was reported that the user interface holder was broken off.There was no patient harm.Manufacturer's ref #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERVO-I
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer (Section G)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer Contact
sara soderholm roman
roentgenvagen 2
solna 
MDR Report Key10850932
MDR Text Key216859236
Report Number8010042-2020-00996
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
PMA/PMN Number
K123149
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSERVO-I
Device Catalogue Number6487800
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/06/2020
Initial Date FDA Received11/17/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/11/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-