• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE, INCORPORATED NUVASIVE RELINE SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE, INCORPORATED NUVASIVE RELINE SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number 11455040
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/22/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product was returned for evaluation.Review of the returned rods found a complete lack of final lock screw torque down contact markings, which indicates the 90-in lbs of required torque was not applied.No additional investigation required.Labeling review: "potential adverse events and complications: as with any major surgical procedures, there are risks involved in orthopedic surgery.Rarely, some complications may be fatal.Potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s)." "warnings, cautions and precautions: the implantation of spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.These devices can break when subjected to the increased load associated with delayed union or nonunion.If healing is delayed, or does not occur, the implant may eventually loosen, bend, or break.Loads on the device produced by load bearing and by the patient¿s activity level will dictate the longevity of the implant.All lock screws should be final-tightened with the counter-torque and torque t-handle.Do not final-tighten through compression instruments (e.G.C/d rack and figure 8 compressor) in the set, as the rod may not be able to normalize to the tulip.Care should be taken to insure that all components are ideally fixated prior to closure." "post-operative warnings: damage to the weight-bearing structures can give rise to loosening of the components, dislocation, and migration, as well as to other complications.".
 
Event or Problem Description
On (b)(6) 2020, a patient underwent a spinal surgery at l4/5.Same day post-operative x-rays showed no evidence of any malfunctions.Three weeks post-op x-rays revealed a left side rod migration.On (b)(6) 2020, a revision surgery was performed where both rods, and tulips were replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE RELINE SYSTEM
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
NUVASIVE, INCORPORATED
7475 lusk boulevard
san diego CA 92121
MDR Report Key10853271
Report Number2031966-2020-00206
Device Sequence Number4169057
Product Code NKB
UDI-Device Identifier00887517591029
UDI-Public887517591029
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K143684
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number11455040
Device Lot NumberCT121831
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2020
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/22/2020
Initial Report FDA Received Date11/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
13550000
-
-