If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device had vibration.Therefore, the reported condition that the device had an undetermined malfunction was confirmed.The assignable root cause of this condition was determined to be traced to component failure due to normal wear.Udi: (b)(4).
|
|
It was reported from switzerland that during service and evaluation, it was determined that the attachment device had missing labeling, the bearing was worn, and the device had vibration and component damage.It was further determined that the device failed pretest for labeling assessment, nose cone verification, and vibration verification.It was noted in the service order that the device had an undetermined malfunction.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2020.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|