• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET MICROFIXATION TMJ SYSTEM LEFT STANDARD MANDIBULAR COMPONENT 50MM / 9 HOLE; JOINT, TEMPOROMANDIBULAR, IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET MICROFIXATION TMJ SYSTEM LEFT STANDARD MANDIBULAR COMPONENT 50MM / 9 HOLE; JOINT, TEMPOROMANDIBULAR, IMPLANT Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Purulent Discharge (1812); Unspecified Infection (1930)
Date of Event 07/08/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet (b)(4).The device will not be returned for analysis as it was discarded; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001032347-2020-00570.Medical products: tmj system left standard mandibular component 50mm / 9 hole, part# 24-6551, lot# 829370d; tmj system left fossa component, medium, part# 24-6561, lot# 863480c; unknown screws, part# ni, lot# ni.The user facility is foreign; therefore, a facility medwatch report will not be available.Report source¿ (b)(6).
 
Event or Problem Description
It was reported the patient underwent a revision to remove temporomandibular joint implants on the left side nine (9) months following implantation due to purulent infection in the ear canal.It is planned that the patient will receive a custom temporomandibular joint implant on the left side in (b)(6) 2021.No additional patient consequences have been reported.
 
Additional Manufacturer Narrative
During the investigation process a review of the sterile certifications were reviewed and found to be conforming with no applicable deviations.Devices were verified to have gone through acceptable sterilization process following iso/aami/astm & eu published guidelines.There are multiple factors that may contribute to an infection that are outside the control of zimmer biomet, such as external factors, i.E.Hospital/surgical environment, provider related risk factors, and/or patient comorbidities/risk factors.As there are no indications of a product or process issues identified affecting implant safety or effectiveness and the reported infection occurred > 90 days, implanted products are not identified as the source or contributing to the reported infection and can be excluded as a possible source.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TMJ SYSTEM LEFT STANDARD MANDIBULAR COMPONENT 50MM / 9 HOLE
Common Device Name
JOINT, TEMPOROMANDIBULAR, IMPLANT
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
MDR Report Key10856513
Report Number0001032347-2020-00569
Device Sequence Number13599020
Product Code LZD
UDI-Device Identifier00841036060329
UDI-Public00841036060329
Combination Product (Y/N)N
Initial Reporter CountryFI
PMA/510(K) Number
P020016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/05/2023
Device Model NumberN/A
Device Catalogue Number24-6551
Device Lot Number829370D
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/02/2020
Supplement Date Received by Manufacturer06/21/2021
Initial Report FDA Received Date11/17/2020
Supplement Report FDA Received Date06/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-