• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR TREK CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012274-20
Medical Device Problem Codes Off-Label Use (1494); Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/28/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the trek rx dilatation catheter was prepped per instructions for use (ifu) without issue.There was no difficulty removing the proactive sheath.The device advanced to the left anterior tibial, moderately calcified lesion without unusual resistance.Upon initial inflation at 6 atmospheres, the balloon burst.Device aspirated was performed and the device was removed without further issues.Reportedly, the aspiration was not performed as a precaution and there was no air embolus.Another trek was successfully used in replacement.There were no adverse patient effects and there was no clinically significant delay.No additional information was provided regarding this issue.
 
Additional Manufacturer Narrative
Visual, functional and scanning electron microscopy (sem) inspections/analysis were performed on the returned device.The reported balloon rupture was not confirmed; however, the noted longitudinal tear in the inner member and balloon is likely what the account perceived as the reported balloon rupture.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined a potential product quality issue based on the noted longitudinal tear in the inner member and balloon which is likely what the account perceived as a reported balloon rupture.The issue is being addressed per internal operating procedures.Abbott vascular will continue to trend the performance of these devices.Na.
 
Event or Problem Description
Subsequent to the previous medwatch report filed, the additional information was reported: the balloon burst at 6 atmospheres.There was no device aspiration performed due to this issue and no adverse patient effect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10856573
Report Number2024168-2020-09633
Device Sequence Number14250394
Product Code LOX
UDI-Device Identifier08717648138447
UDI-Public08717648138447
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K103110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2022
Device Model Number1012274-20
Device Catalogue Number1012274-20
Device Lot Number00110G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2020
Initial Date Received by Manufacturer 10/28/2020
Supplement Date Received by Manufacturer02/09/2021
Initial Report FDA Received Date11/17/2020
Supplement Report FDA Received Date02/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age74 YR
Patient Weight89
-
-