Model Number RBY4C1650-B
Medical Device Problem Codes Physical Resistance/Sticking (4012); Premature Separation (4045)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/23/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the iliac vein using ruby coils and a lantern delivery microcatheter (lantern).During the procedure, while advancing a ruby coil through the lantern, the physician experienced resistance and decided to retract the ruby coil.However, while attempting to retract and re-sheath the ruby coil, it unintentionally detached in the patient¿s body.Therefore, the pusher assembly of the ruby coil was removed from the lantern and the ruby coil was snared.The procedure was completed using another ruby coil and the same lantern.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Please note that the device associated with this complaint was expected to be returned; however, additional information received indicated that the device was disposed of and is no longer available for return.Therefore, without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key10857164
Report Number3005168196-2020-02029
Device Sequence Number2420104
Product Code HCG
UDI-Device Identifier00814548018904
UDI-Public00814548018904
Combination Product (Y/N)Y
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K173614
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/23/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRBY4C1650-B
Device Catalogue NumberRBY4C1650
Device Lot NumberF95508
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 10/23/2020
Supplement Date Received by Manufacturer12/15/2020
Initial Report FDA Received Date11/17/2020
Supplement Report FDA Received Date02/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age85 YR
Date Report Sent to FDA01/01/2005
-
-