| Model Number |
RBY4C1650-B |
| Medical Device Problem Codes |
Physical Resistance/Sticking (4012); Premature Separation (4045)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/23/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure in the iliac vein using ruby coils and a lantern delivery microcatheter (lantern).During the procedure, while advancing a ruby coil through the lantern, the physician experienced resistance and decided to retract the ruby coil.However, while attempting to retract and re-sheath the ruby coil, it unintentionally detached in the patient¿s body.Therefore, the pusher assembly of the ruby coil was removed from the lantern and the ruby coil was snared.The procedure was completed using another ruby coil and the same lantern.There was no report of an adverse effect to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Please note that the device associated with this complaint was expected to be returned; however, additional information received indicated that the device was disposed of and is no longer available for return.Therefore, without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
| |
|