• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. UNIV COCR 14/16 FEM HEAD 32MM +8; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. UNIV COCR 14/16 FEM HEAD 32MM +8; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 129966
Medical Device Problem Codes Use of Device Problem (1670); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Failure of Implant (1924); Injury (2348); Joint Dislocation (2374)
Date of Event 10/25/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after a tha had been performed, the femoral head disassociated with the stem.A revision surgery was necessary; the sales representative attended the revision, where the surgeon noticed severe damage to the stems trunnion and had to remove the stem in order to proceed with a full revision.When the femoral head was retrieved, it was confirmed that the wrong taper was originally used and was ultimately the cause of the disassociation/damage.The patient current status of health is unknown.The stem that had been initially implanted was a size 0 standard collared polartstem, and instead of the correct 12/14 tapered femoral head, a 32mm +8 large tapered 14-16 head was accidentally picked by the nursing staff and implanted into the patient.The sales representative had not been present during this surgery, as covid-19 restrictions prohibited vendors from entering the hospital at the time of surgery.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, based on the surgeon¿s statement, ¿when the femoral head was retrieved, it was confirmed that the wrong taper was originally used and was ultimately the cause of the dissociation/damage;¿ we cannot rule out a user vs procedural error as the likely cause of the reported event and not the performance of the smith and nephew device.Due to the limited documentation provided, the impact to the patient beyond the incorrect product implantation, disassociation and revision cannot be determined.Should any additional medical information be provided this complaint will be re-assessed.Therefore, no further clinical/medical assessment is warranted at this time.A review of complaint history did not reveal additional complaints for the listed device.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include but are not limited to procedural variance or user error.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIV COCR 14/16 FEM HEAD 32MM +8
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10857510
Report Number1020279-2020-06552
Device Sequence Number11413385
Product Code JDI
UDI-Device Identifier03596010087515
UDI-Public03596010087515
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K870128
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number129966
Device Catalogue Number129966
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/26/2020
Supplement Date Received by Manufacturer08/19/2021
Initial Report FDA Received Date11/17/2020
Supplement Report FDA Received Date08/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-