| Model Number |
D-AVSE-DF20-V1525 |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Thrombus (2101)
|
| Date of Event |
10/21/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results, method and conclusion codes along with investigation results will be provided in the final report.
|
| |
|
Event or Problem Description
|
|
Related manufacturer report number: 3008452825-2020-00604.After a paroxysmal atrial fibrillation ablation procedure, a blood clot was noted to be attached to the electrode 7.An echocardiogram was performed after the procedure, and no floating substance/material like a blood clot was found in the left atrium and there were no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
One 20 pole, bi-directional, curve d-f, sensor enabled, advisor vl circular mapping catheter loop was received for evaluation.The distal shaft was cut and the loop was returned.The catheter was not returned.A blood like substance was noted on electrodes 7,8, 13 and 14.Functional testing was not possible due to the condition of the returned catheter.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the blood like substance remains unknown.
|
| |
|