| Catalog Number |
321-55-42 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/03/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pending evaluation.
|
| |
|
Event or Problem Description
|
|
As reported, during a procedure on a male patient, as the reamer was being taken off modular driver after reaming was finished the tip broke into 2 pieces.No parts or pieces fell into the wound site.There was no surgical delay.Patient was last known to be in stable condition following the event.Device to be returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h10: (h3) the broken device reported was likely the result of applying a counter clockwise resistance force to the instrument greater than yield strength of the material and subsequent sterilization cycles, which led to crack initiation, propagation, and ultimate failure.
|
| |
|