| Model Number |
TRL-260 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
10/21/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side rupture.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified the device was broken shell and missing shell.A weight test of the device was verified and the device was underweight.A microscopic analysis was performed which identified broken sharp.A dimension measurement in the shell was performed which identified the thickness within specification.Based on the device analysis, the final assessment is a sharp edge broken in the side anterior assessed as unidentified (tear) opening and missing shell assessed as inconclusive.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side rupture.Device has been explanted.
|
| |
|