• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE (127 DEG) SIZE 2.5 ACCOLADE (127 DEG) SIZE 2.5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE (127 DEG) SIZE 2.5 ACCOLADE (127 DEG) SIZE 2.5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 6021-2530
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Pain (1994)
Date of Event 10/27/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.Not returned.
 
Event or Problem Description
It was reported that the patient's right hip was revised due to pain and elevated cocr levels.Intra-operatively, the following were noted: gray tissue, bone resorbtion, and black debris inside the head/ trunnion interface.The surgeon cleaned off the trunnion of the in-situ stem and revised a cocr femoral head and liner to a biolox ceramic head with sleeve, mdm metal liner, and adm/ mdm poly insert.No allegations were made against the revised liner.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE (127 DEG) SIZE 2.5 ACCOLADE (127 DEG) SIZE 2.5
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key10860945
Report Number0002249697-2020-02489
Device Sequence Number2862701
Product Code LPH
UDI-Device Identifier04546540510747
UDI-Public04546540510747
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2017
Device Model Number6021-2530
Device Catalogue Number6021-2530
Device Lot Number40746002
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/27/2020
Initial Report FDA Received Date11/18/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age71 YR
-
-