| Model Number |
IPN000099 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/05/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn#: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that an swg (0.025inch) did not go through the catheter.As a result, the user opened a new kit, and the swg successfully went through the new catheter and the procedure was completed without problem.There was no report of delay in therapy.There was no report of patient complications, serious injury or death.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).Teleflex received the device for investigation.The reported complaint that the "swg did not go through the catheter " is confirmed.Upon return, the injection lumen failed aspiration and flush testing due to a blood clot.The blood clot was cleared from the injection lumen after the guidewire test and another attempt of aspirating and flushing was successfully completed.The root cause of the complaint is undetermined.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.Teleflex assessed the risk for the reported complaint.There are no new or revived risk.This will be monitored for any developing trends.
|
| |
|
Event or Problem Description
|
|
It was reported that an swg (0.025inch) did not go through the catheter.As a result, the user opened a new kit, and the swg successfully went through the new catheter and the procedure was completed without problem.There was no report of delay in therapy.There was no report of patient complications, serious injury or death.
|
| |
|
Search Alerts/Recalls
|