Model Number N/A
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/21/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A review of the manufacturing documentation associated with lot 17955919 presented no issues during the manufacturing process that can be related to the reported event.Additional information is pending and will be submitted within 30 days upon receipt.
 
Event or Problem Description
When the surgeon took out the 5f 100cm straight (st) sidewinder simmons technique (sim1) super torque angiographic catheter from the patient, the tip disengaged from the catheter.There was no reported patient injury.The device will be returned for analysis.
 
Additional Manufacturer Narrative
Complaint conclusion: when the surgeon took out the 5f 100cm straight (st) sidewinder simmons technique (sim1) super torque angiographic catheter from the patient, the tip disengaged from the catheter.There was no reported patient injury.Additional procedural details were requested but not provided.The device was not returned for analysis.A product history record (phr) review of lot 17955919 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.Based on the information provided, it is not possible to draw a clinical conclusion between the device and the reported event.Without the return of the device for analysis, the reported customer event ¿brite tip/distal tip - catheters- separated¿ could not be confirmed and the exact root cause could not be determined.Procedural/handling factors or vessel characteristics (although not provided) could possibly contribute to this type of event.Per the instructions for use (ifu), which is not intended as a mitigation of risk, ¿exercise care when removing guidewires from multiple-curve catheters.To prevent kinking of 5f (1.65 mm) and smaller angiographic catheters; straighten the pigtail catheter tip only with a diagnostic guidewire or, if applicable, with a tip straightener.Do not straighten by hand.Use a guidewire when introducing the catheter through the catheter sheath introducer (csi) and into the left ventricle.Treat all 4f catheters and smaller french sizes with ultimate care.The performance of these products may be impaired if not properly and cautiously handled during unpacking and preparation.Procedures requiring percutaneous catheter introduction should not be attempted by physicians unfamiliar with the possible complications.Complications may occur at any time during or after the procedure.¿ neither the phr review nor the information available suggests that the event could be related to the manufacturing process of the unit.Therefore, no corrective or preventive actions will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CATH ST 5F SIM 1 100CM
Common Device Name
CATHETER, INTRAVASCULAR, DIAGNOSTIC
MDR Report Key10861451
Report Number9616099-2020-04069
Device Sequence Number2559249
Product Code DQO
UDI-Device Identifier10705032011648
UDI-Public10705032011648
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K915836
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2023
Device Model NumberN/A
Device Catalogue Number532501
Device Lot Number17955919
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/23/2020
Supplement Date Received by Manufacturer11/25/2020
Initial Report FDA Received Date11/18/2020
Supplement Report FDA Received Date12/01/2020
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/30/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
-
-