• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AZIYO BIOLOGICS INC. CANGAROO ENVELOPE; MESH, SURGICAL - FTM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AZIYO BIOLOGICS INC. CANGAROO ENVELOPE; MESH, SURGICAL - FTM Back to Search Results
Model Number CMCV-009-MED
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/07/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing review of the cangaroo envelope device history record for the reported lot shows that all packaged and labeled units were quality released to distribution on 04/17/2020 having met all internal qc acceptance requirements.All sterilization processing records and bioburden testing indicate a successful sterilization cycle and passing results of product lal and sterility samples allowing the subassembly lot to be released for further packaging and labeling having met all criteria for release.There were no non-conformances during manufacturing or sterilization potentially impacting the final acceptance of this manufacturing lot.In lieu of a request for the oem supplier dhr review, it is noted that aziyo processes the non-sterile envelope materials including cutting, suturing, packaging and sterilizing the product.The reason for the explantation of the aziyo device is was related to a tuning issue with the implanted cied and device leads resulting in high pacing thresholds which would potentially decrease battery life.A replacement device and cangaroo envelope were implanted following removal.Device tuning issues are a recognized procedural complication associated with cied implantation, and not associated with usage of the cangaroo envelope.Should aziyo receive any additional details related to this event, a supplemental report will be filed.
 
Event or Problem Description
It was reported by aziyo biologics' business partner boston scientific, that a cangaroo envelope (model cmcv-009-med lot #m20d1137) was used with a boston scientific device (model # / serial # unknown) and implanted on (b)(6) 2020.On (b)(6) 2020 the device and cangaroo pouch were explanted due to a lead revision procedure required due to high pacing thresholds.A new device and cangaroo envelope were implanted.This issue was reported to aziyo biologics by boston scientific on 10/22/2020 and further information received on 11/18/2020.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CANGAROO ENVELOPE
Common Device Name
MESH, SURGICAL - FTM
Manufacturer (Section D)
AZIYO BIOLOGICS INC.
1100 old ellis road
ste 1200
roswell GA 30076
MDR Report Key10861895
Report Number3005619880-2020-00095
Device Sequence Number3949473
Product Code FTM
UDI-Device Identifier00859389005034
UDI-Public00859389005034
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K192616
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other,use
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/10/2022
Device Model NumberCMCV-009-MED
Device Lot NumberM20D1137
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/22/2020
Initial Report FDA Received Date11/18/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BSC CIED (MODEL# / SERIAL# - UNKNOWN)
Outcome Attributed to Adverse Event Required Intervention;
-
-