• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problem Noise, Audible (3273)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 10/26/2020
Event Type  malfunction  
Manufacturer Narrative
A getinge service territory manager (stm) was dispatched to investigate.The stm reported that the repair is complex and ongoing, due to the fact that it is based on a saline spill.The stm has to order additional parts to complete the service.Once the repair is completed, additional information will be submitted.
 
Event Description
It was reported that the cardiosave intra-aortic balloon pump (iabp) was making a beeping noise at start up.It is unknown if there was any patient harm/injury; however there was no adverse event reported.
 
Event Description
It was reported that at start up the cardiosave intra-aortic balloon pump (iabp) was making a beeping noise.There was no patient involvement, and no adverse event reported.
 
Manufacturer Narrative
The getinge service territory manager (stm) that had reported that the repair was complex and ongoing, due to saline spill, reported that significant fluid intrusion were on internal components.Multiple circuit boards were affected.During the repair process, however fluid intrusion caused the immediate failure of each new board, resulting in the need to replaced some boards more than once.At this time the customer has elected to halt all repairs of this iabp.No further work was performed by the stm, and the iabp was not cleared for clinical use.It was also reported that the customer retired this iabp from clinical service.
 
Manufacturer Narrative
Analysis of production: (3331/213) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109/213) the review of the historical data indicates that no other similar complaint was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110/213) the overall 12 month product complaint trend data for the period nov 2019 through oct 2020 was reviewed.There were no triggers identified for the review period.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
brian schaeffer
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key10862222
MDR Text Key217402613
Report Number2249723-2020-01952
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 02/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/18/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/28/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/20/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-