Model Number 0998-00-0800-53 |
Device Problem
Noise, Audible (3273)
|
Patient Problems
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/26/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge service territory manager (stm) was dispatched to investigate.The stm reported that the repair is complex and ongoing, due to the fact that it is based on a saline spill.The stm has to order additional parts to complete the service.Once the repair is completed, additional information will be submitted.
|
|
Event Description
|
It was reported that the cardiosave intra-aortic balloon pump (iabp) was making a beeping noise at start up.It is unknown if there was any patient harm/injury; however there was no adverse event reported.
|
|
Event Description
|
It was reported that at start up the cardiosave intra-aortic balloon pump (iabp) was making a beeping noise.There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
The getinge service territory manager (stm) that had reported that the repair was complex and ongoing, due to saline spill, reported that significant fluid intrusion were on internal components.Multiple circuit boards were affected.During the repair process, however fluid intrusion caused the immediate failure of each new board, resulting in the need to replaced some boards more than once.At this time the customer has elected to halt all repairs of this iabp.No further work was performed by the stm, and the iabp was not cleared for clinical use.It was also reported that the customer retired this iabp from clinical service.
|
|
Manufacturer Narrative
|
Analysis of production: (3331/213) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109/213) the review of the historical data indicates that no other similar complaint was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110/213) the overall 12 month product complaint trend data for the period nov 2019 through oct 2020 was reviewed.There were no triggers identified for the review period.
|
|
Event Description
|
N/a.
|
|