| Model Number |
15 |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/06/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Physio-control evaluated the customer's device and was unable to duplicate the reported issue however, the customer did not provide the batteries that were in use with the device at the time of the event.Physio-control continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device unexpectedly shut off.There was no patient use reported with the event.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device unexpectedly shut off.There was no patient use reported with the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Physio-control evaluated the customer's device and duplicated the reported issue.Physio replaced the device's corroded battery contact pcb assembly to resolve the reported issue.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.
|
| |
|