• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH GLENOSPHERE - 32MM DIA X 2MM THK CONCENTRIC; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH GLENOSPHERE - 32MM DIA X 2MM THK CONCENTRIC; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Model Number 5573-C-3202
Medical Device Problem Codes Material Erosion (1214); Device Damaged by Another Device (2915)
Health Effect - Clinical Codes Failure of Implant (1924); Implant Pain (4561)
Date of Event 10/26/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Upon completion of investigation, additional information will be provided in a supplemental report.
 
Event or Problem Description
It was reported that the patient's right reverse shoulder was revised after patient complaint of pain and x-rays showing a peripheral screw loose in the joint.Intra-operatively, it was noted that the peripheral screws had 'windshield wipered' in the joint, etching the glenosphere and passing through the baseplate into the joint.The baseplate, glenosphere and center screw came out all in one piece.
 
Additional Manufacturer Narrative
The reported event could partially be confirmed, based on the state of the implants returned.The received images revealed that the glenosphere presents two black traces on the metal, caused by the peripheral screws backing out.The joint surface is smooth and does not show any major scratches or deformations.More detailed information about the complaint event such as the x-rays must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If any further information is provided, the investigation report will be updated.
 
Event or Problem Description
It was reported that the patient's right reverse shoulder was revised after patient complaint of pain and x-rays showing a peripheral screw loose in the joint.Intra-operatively, it was noted that the peripheral screws had 'windshield wipered' in the joint, etching the glenosphere and passing through the baseplate into the joint.The baseplate, glenosphere and center screw came out all in one piece.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLENOSPHERE - 32MM DIA X 2MM THK CONCENTRIC
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
MDR Report Key10864580
Report Number0008031020-2020-02483
Device Sequence Number11478877
Product Code KWS
UDI-Device Identifier07613327098877
UDI-Public07613327098877
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K130895
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/30/2020
Device Model Number5573-C-3202
Device Catalogue Number5573-C-3202
Device Lot Number738122
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2020
Initial Date Received by Manufacturer 10/26/2020
Supplement Date Received by Manufacturer01/25/2021
Initial Report FDA Received Date11/18/2020
Supplement Report FDA Received Date02/10/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
-
-