• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER Back to Search Results
Model Number CV-180
Device Problem Poor Quality Image (1408)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/20/2020
Event Type  malfunction  
Manufacturer Narrative
An olympus field service engineer (fse) was dispatched to the user facility to evaluate/repair the suspect device.The fse was able to duplicate reported problem and assigned the cause to a defective videoscope cable.
 
Event Description
A user facility reported to olympus that the image appeared yellow.The problem, as reported to olympus, was found prior to performing a procedure.The intended procedure was completed using the same device without a delay.There was no patient injury or harm associated with the reported problem.
 
Manufacturer Narrative
This supplemental report is submitted to provide the results of the legal manufacturer¿s investigation and device history record (dhr) review.The dhr for the subject device was reviewed and it was verified the device was manufactured in accordance with documented specifications.The legal manufacturer performed an investigation.A conclusive root cause was not identified.The legal manufacturer determined the likely cause was failure of the video scope cable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II VIDEO SYSTEM CENTER
Type of Device
VIDEO SYSTEM CENTER
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10864639
MDR Text Key225187250
Report Number8010047-2020-09102
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170215513
UDI-Public04953170215513
Combination Product (y/n)N
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 01/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCV-180
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/21/2020
Initial Date FDA Received11/18/2020
Supplement Dates Manufacturer Received12/18/2020
Supplement Dates FDA Received01/14/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-