• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH BASEPLATE; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH BASEPLATE; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Model Number 5572-2800
Medical Device Problem Code Separation Problem (4043)
Health Effect - Clinical Codes Failure of Implant (1924); Implant Pain (4561)
Date of Event 10/26/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Upon completion of investigation, additional information will be provided in a supplemental report.
 
Event or Problem Description
It was reported that the patient's right reverse shoulder was revised after patient complaint of pain and x-rays showing a peripheral screw loose in the joint.Intra-operatively, it was noted that the peripheral screws had 'windshield wipered' in the joint, etching the glenosphere and passing through the baseplate into the joint.The baseplate, glenosphere and center screw came out all in one piece.
 
Event or Problem Description
It was reported that the patient's right reverse shoulder was revised after patient complaint of pain and x-rays showing a peripheral screw loose in the joint.Intra-operatively, it was noted that the peripheral screws had 'windshield wipered' in the joint, etching the glenosphere and passing through the baseplate into the joint.The baseplate, glenosphere and center screw came out all in one piece.
 
Additional Manufacturer Narrative
The reported event could partially be confirmed, based on the state of the implants returned.The pictures provided revealed that the screw holes where the two peripheral screws were inserted have expended, in such a way that the material around it is very damaged.On one side of the baseplate, the screw hole has been damaged so severely that there is hardly any material left on one border.The deformation also makes the determination of the lot number for the baseplate impossible since the material around it disappeared.More detailed information about the complaint event such as the x-rays must be available in order to determine the root cause of the complaint event.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BASEPLATE
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
MDR Report Key10864670
Report Number0008031020-2020-02484
Device Sequence Number11478878
Product Code KWS
UDI-Device Identifier07613327098709
UDI-Public07613327098709
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K130895
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number5572-2800
Device Catalogue Number5572-2800
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2020
Initial Date Received by Manufacturer 10/26/2020
Supplement Date Received by Manufacturer01/05/2021
Initial Report FDA Received Date11/18/2020
Supplement Report FDA Received Date02/01/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
-
-