Model Number 20E
Device Problem Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/03/2020
Event Type  malfunction  
Manufacturer Narrative
Physio-control evaluated the customer's device but was unable to duplicate or verify the reported power issue. After observing proper device operation through functional and performance testing the unit was returned to the customer for use. The cause of the reported issue could not be determined.
 
Event Description
The customer contacted physio-control to report that their device logged a critical event code. The event code is indicative of a device issue that could result in a partial loss of defibrillator output energy due to a loss of the negative or positive portion of the biphasic output waveform. The defibrillator output energy could be reduced by up to approximately 20% from the selected energy level. There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameLIFEPAK(R) 20E DEFIBRILLATOR/MONITOR
Type of DeviceAUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key10864993
MDR Text Key217192485
Report Number0003015876-2020-01746
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130454
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation
Type of Report Initial
Report Date 11/18/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/18/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number20E
Device Catalogue Number99507-000106
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2020
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received11/03/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/19/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-