It was reported that upon checking the camera system before an arthroscopy procedure, it was noticed that the picture was not clear, it was black when looking through the videoarthroscope which appears damaged.The procedure was completed using a competitor device with a delay shorter than 30 minutes.No other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
H6: the reported device, intended for use in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.No product identification information was provided and thus a manufacturing record review could not be conducted.A complaint history review concluded this was a repeat issue.A review of risk management files found that the reported failure was documented appropriately.A review of the instructions for use found: prior to each use, inspect the device to ensure it is functioning properly and not damaged.Do not use a damaged device.To avoid fogging during surgery, the rear portion of the scope must be entirely free of moisture before attachment to either the camera head or camera coupler.Caution must be taken to dry-out mating portions of both the scope and camera head; otherwise trapped moisture may condense on glass surfaces, blurring the image.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
|