• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER DENTAL IMPL TAPERED SCR-V HA 3.7 MM 3.5MM 16MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL IMPL TAPERED SCR-V HA 3.7 MM 3.5MM 16MM; DENTAL IMPLANT Back to Search Results
Model Number TSVH16
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Information (3190)
Date of Event 07/01/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Weight unknown / not provided.Premarket identification pma/510(k) number k011028 and k013227.A summary investigation has been completed for lack of primary stability events that do not allege a deficiency with the implant and identified that the reported event is likely due to biological factors which have an adverse effect on implant stability or is related to surgical technique and insertion torque.Due to a wide range of external factors (non-design/ non-manufacturing related), identifying a definitive root cause is generally not possible.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
Doctor reported the implant in tooth location 11 had lack of primary stability and was removed.Doctor did not place another implant in the same surgery.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPL TAPERED SCR-V HA 3.7 MM 3.5MM 16MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
MDR Report Key10867628
Report Number0002023141-2020-02009
Device Sequence Number2601816
Product Code DZE
UDI-Device Identifier00889024019591
UDI-Public(01)00889024019591(17)230228(10)63908029(241)TSVH16
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K061410
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2023
Device Model NumberTSVH16
Device Catalogue NumberTSVH16
Device Lot Number63908029
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/27/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 10/27/2020
Supplement Date Received by Manufacturer02/03/2021
Initial Report FDA Received Date11/19/2020
Supplement Report FDA Received Date02/11/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
-
-