• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC ROD INTRODUCTION PLIERS FOR SIDE-OPENING IMPLANTS; PLIERS,SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC ROD INTRODUCTION PLIERS FOR SIDE-OPENING IMPLANTS; PLIERS,SURGICAL Back to Search Results
Model Number 388.50
Medical Device Problem Code Compatibility Problem (2960)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/22/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2020, the patient underwent a scoliosis t10-l5 fusion procedure.Upon inspection, during setup, it was noticed that the handles of two (2) socket wrench with straight handle (with silicone handle) had a residue of white powder in the cannula resulting in the entire sterile field being contaminated.During the surgery, two (2) rod introduction pliers were used and ended up getting stuck inside a hook/screw holder and the handle of a socket wrench with straight handle disintegrated upon routine tightening.The procedure was successfully completed with a thirty (30) minute surgical delay.The patient's status is unknown.Concomitant device reported: unknown nut (part#: unknown, lot#: unknown, quantity: 1) this report is for one (1) rod introduction pliers for side-opening implants.This is report 2 of 3 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Service and repair history: the previous service event for part number 388.50 with lot number(s) t124099 has been reviewed.The customer called in a service request for this item on oct 22, 2020 for handles of two (2) socket wrench with straight handle (with silicone handle) had a residue of white powder in the cannula.The item was previously returned for service on (b)(6) 2018 due to tube burred at slot.The previous service condition of tube burred at slot is not relevant to the current complained issue of handles of two (2) socket wrench with straight handle (with silicone handle) had a residue of white powder in the cannula.The manufacture date of this item is dec 29, 2015.The service history review is unconfirmed.Service and repair evaluation: the repair technician reported the collet and tube were loose.Loose component is the reason for repair.The cause of the issue is unknown.The following parts were replaced: tube, nut, collet, and all applicable components.The item was repaired per the inspection sheet, passed synthes final inspection on nov 12, 2020 and will be returned to the customer upon completion of the service and repair process.Attached service record router completed through operation 30.Finalized service record will be archived in tungsten document management system.The evaluation was confirmed.The device was deemed serviceable and will be returned to the customer, no design or manufacturing issues were identified therefore it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROD INTRODUCTION PLIERS FOR SIDE-OPENING IMPLANTS
Common Device Name
PLIERS,SURGICAL
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key10868354
Report Number2939274-2020-05223
Device Sequence Number3949250
Product Code HTC
UDI-Device Identifier10705034773292
UDI-Public(01)10705034773292
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number388.50
Device Catalogue Number388.50
Device Lot NumberT124099
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2020
Initial Date Received by Manufacturer 10/22/2020
Supplement Date Received by Manufacturer11/19/2020
12/16/2020
Initial Report FDA Received Date11/19/2020
Supplement Report FDA Received Date12/15/2020
12/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
HOOK OR SCREW HOLDER; ROD INTRODUCTION PLIERS FOR SIDE-OPENING IMPLANTS; SOCKET WRENCH WITH STRAIGHT HANDLE-11MM WIDTH; SOCKET WRENCH WITH STRAIGHT HANDLE-11MM WIDTH; SOCKET WRENCH WITH STRAIGHT HANDLE-11MM WIDTH; UNK - NUTS
-
-